Font Size:
Site Colors:
Accesskey
S - Skip navigation
1 - Home page
4 - Search
Accessibility Statement
Setting New Standards In Electro-Optical Testing

SR-800N Superior Accuracy Blackbody

Download Brochure
SR-800N (2315KB)

CI Systems’ advanced architecture ControlMaster
SR-800N sets a new standard for accuracy and uniformity in blackbody technology.

Temperature measurement and calibration are both performed in the radiation head itself and then transferred digitally to the ControlMaster controller.

The result is accurate, stable, reliable and NIST- traceable.

Temperature is controlled by easily-removable sensors which the user can replace in just minutes.

Our factory-supplied replacement sensors provide an efficient means for recalibrating the system, providing an additional 12 months of service.

Alternatively, customers may recalibrate the blackbody with our optional

CK-800R calibration kit.

Ask for quotation

More information
FeaturesOptionsSpecificationsCertifications
  • Superior accuracy
  • High-uniformity emitting surface
  • Millidegree-Kelvin resolution
  • Wide range of radiation temperatures
  • Able to operate at a wide range of ambient temperatures
  • Configurable temperature resolution
  • Configurable stability window
  • Quick periodic calibration by replacing the removable sensor
  • Dedicated remote control software included
  • Large color LCD display with touch screen user interface
  • Certified to MIL-T-28800D, CE, and FCC
  • NIST-traceable calibration
  • Standard emitter sizes ranging from 2” to 40“. Other sizes are available upon request 

  • Low acoustic noise

  • Compact, portable ControlMaster controller

  • 19” rack-mount kit included

  • Ethernet and RS-232 communication ports

  • Nitrogen inlet for inert atmosphere on all LT models 

  • Ability to control up to four motorized devices (optional) 

  • Dual head (optional)

  • Calibration kit CK-800 available upon request 

 

MODEL: SR-800N-

2A

2D

4A

4D

7A

7D

8A

8D

10A

10D  

12A

12D

14A

 14D

16A

 16D 

20A

 20D

ET - Extended temp.

Absolute temp. range,°C

Differential temp. range,°C

0 to 175

-25 to 150

0 to 125

-25 to 100

0 to 125

-25 to 100

LT - Low temp. (4)

Absolute temp. range,°C

Differential temp. range,°C

---

---

  

-40 to 150

-65 to 125 

---

---

---

---

---

---

WTR -
Wide temp. range

Absolute temp. range,°C

Differential temp. range,°C 

 

---

---

  

-20 to 150

-45 to 125

---

---

---

---

---

---

---

---

---

---

---

---

HE - High Emissivity

0.99 ± 0.01

Environmental Chamber

   
 

CH-STD

Chamber temp. °C

Abs. temp. range °C

Diff. temp. range °C

-30 to 70

-40 to 80

-10 to 40

 

CH-ET

 

Chamber temp. ° C

Abs. temp. range °C

Diff. temp. range °C

-40 to 80

-40 to 150

-20 to 125

-40 to 80

-40 to 150

-20 to 100

-40 to 80

-40 to 150

-15 to 100

 

CH-LT (4)

 Chamber temp. °C

Abs. temp. range °C

Diff. temp. range °C

 

---

---

---

-40 to 80

-40 to 150

-20 to 125

---

---

---

---

---

---

---

---

---

Differential Accuracy °C

0.020 @ -20 < T ambient < 80

0.040 @ T ambient < -20

Stability,°C

0.005 @ ΔT < 10

0.010 @ ΔT > 10

Dual head

Absolute temp. range,°C

Differential temp. range,°C

0 to 125

-25 to 100

 

---

---

 

---

---

 

---

---

 

---

---

 

---

---

---

---

GPIB optional

 

 See the brochure for more information 

SR-800 Model

2A

 2D

4A

 4D

7A

 7D

8A

8D

10A

 10D  

12A

12D

14A

 14D

16A

 16D 

20A

 20D

40A

 40D

Emitter size

2” dia.

4”x4”

7”x7”

8”x8”

10”x10” 

12”x12”

14”x14”

16”x16”  

20”x20”

40”x40” 

Absolute temp. range, °C

0ºC to

125ºC

 0ºC to

100ºC 

10ºC to

80ºC

Differential temp. range,°C

-25ºCto

100ºC

-25ºC to

75ºC

-15ºC to

55ºC

Uniformity, °C (2)

± 0.005ºC

± 0.010

± 0.015ºC

± 0.030

Set point resolution, °C

 

0.001

Temp. accu- racy:

 

Absolute, °C (3)

0.015 @T< 0, or 0.007 @ 0 < T < 50, or 0.015 @ T > 50

Differential, °C (3)

 

0.008 @ ǀΔTǀ ≤ 25 or 0.015 @ ǀΔTǀ > 25

Stability (at absolute mode), °C

±0.003 @ ǀΔTǀ ≤ 10 or ± 0.008 @ ǀΔTǀ > 10

Emissivity

0.98± 0.01

Settling time, sec (4)

30

Power consumption

200

200

600

1000

1000

1200

2400

2800

3500

7000

 

Size & weigh

 

 
 
 

 

 

 

Blackbody Head

   

Height, mm

Ø 65

161

224

350

350

357

410

500

570

1280

Width, mm

201

265

305

305

415

480

550

640

 

760

Depth, mm

95

190

225

 170

170

230

230

230

230

1600

Weight, kg

 10

15 

 20  25

36 

 42  65  450

Controller

 

HxWxD, mm

342x310x133 (3U)

450x600x175 (4U)

 

Weight, kg

 

10     

 20  20

25

60

Operating voltage

95 to 240 VAC. 50/60 Hz          

Operating temp. Head, °C  -20 to +70         

Operating temp., Controller, °C

0 to 50

Storage temp., °C  -20 to +70         

 

Notes:

(1) All values are valid at an ambient temperature of 22°C and in a non- condensing environment.

(2) Uniformity values are for a ±1°C step from ambient temp @ 80% of the central area. For other temperatures multiply by T.

(3) Accuracy is referenced to a NIST-calibrated CI Systems master sensor.

(4) Settling time is to a deviation of 0.01°C from the required value.

(5) Includes refrigerator (power consumption: 3500 W).

(6) Differential temperature range is limited to absolute temperature range.

(7) Total system uncertainty: 0.02°C @ T < ±25°C, 0.03°C @ T > ±25°C.

(8) Typical yearly drift is 0.02°C.

(9) Specifications are subject to change without notice

(10) See examples of special applications on page 4 

  • NIST-traceable calibration
  • Certified to MIL-T-28800D, CE, and FCC